How to measure an RF Amplifier using a DSA-800 Series Spectrum Analyzer
How to measure an RF Amplifier using a DSA-800 Series Spectrum Analyzer Date:12.20.20110 Solution: This document provides step-by-step instructions on using the Rigol DSA-800 series of Spectrum
How do I measure an RF Bandpass filter
How to measure an RF Bandpass filter using a DSA815TG Spectrum Analyzer Date:12.20.2011 Solution: This document provides step-by-step instructions on using the Rigol DSA-800 series of Spectrum Ana
What about RIGOL battery and portable use? Can you run RIGOL Scopes and Spectrum Analysers from Battery or portable UPS Power Supplies?
Running the DSA815 from a battery (see also comment about other instruments) Although the size and weight of the DSA800's make it attractive for use in locations where there is no pow
RIGOL 3GHz Spectrum Analyser - Applications insights from RIGOL-UK's customers.
1. Simple modulation testing above UHF. First I get straight on with monitoring typical modulation sidebands, I’m at 2,320MHz with DSA1030A-TG3 - a performance spectrum analyser from RIGOL incor
How do I achieve the lowest DANL with my spectrum analyser?
Set your DSA as follows: Attenuator = 0dB; Preamplifier ON; Center Frequency = 10MHz; RBW = 10Hz; Span = 10KHz. This is a typical procedure for all professional analysers.