EMC Precompliance: Testing on a budget
Last Updated: 21/11/2013
EMC Precompliance: Testing on a budget

Date:29.05.2012

Solution: Almost any electronic design slated for commercial use is subject to EMC (Electromagnetic Compatibility) testing. Any company intending to sell  these products into a country must ensure that the product is tested versus specifications set forth by the regulatory body of that country. Here in the US, the FCC specifies rules on EMC testing. CISPR and IEC are also used throughout the world.

To be sold legally, a sample of the electronic product must pass a series of tests. In many cases, companies can self-certify, but they must have detailed reports of the test conditions and data. Many companies choose to have these tests performed by accredited compliance company. This full compliance testing can be very expensive with many labs charging thousands of dollars for a single day of testing. Testing a product for full compliance can also require specialised testing environments and no changes can be made during the testing. Any failures in compliance testing require that the design heads back to Engineering for analysis and possible redesign. This can cause delays in product release and an obvious increase in design costs.

One method to lower the additional costs associated with EMC compliance is to perform EMC testing throughout the design process well before sending the product off for full compliance testing. This pre-compliance testing can be very cost effective and can be tailored to closely match the conditions used for compliance testing. This will increase your confidence in passing compliance the first time through, lower your test costs, and speed your time to market.


Measuring Radiated EMI
The most simple form of pre-compliance measurements for radiated emissions can be performed using a spectrum analyser, like the Rigol DSA-815 (9kHz to 1.5GHz), and near field electric (E) and magnetic (H) probes.  

  Figure 1 and 2: Near field E and H probes.

The most simple test is to configure the DSA to use the peak detector and set the RBW and Span for the area of interest per the regulatory requirements for your device. Then select the proper E or H probe for your design and scan over the surface of the design.

Probe orientation (rotation, distance) is also important to consider. The probes act as an antenna, picking up radiated emissions from seams, openings, traces, and other elements that could be emitting RF. A through scan of all of the circuit elements, connectors, knobs, openings in the case, and seams is crucial.


              Figure 3: Using a near field H probe
        to test a power supply.




For the first pass, configure the spectrum analyser to use the peak detector. This will provide you with a “worst case” reading on the radiated RF and it is the quickest path to determining the problem areas. Larger probes will give you a faster scanning rate, albeit with less spacial resolution.

Once you have a good idea of your problem areas, you can get more detail by implementing a few common techniques. If you can, select a spectrum analyser that has the standard configuration used in full compliance testing. This includes a Quasi-Peak detector mode, EMI filter, and Resolution Bandwidth (RBW) settings that match the full test requirements specified for your product.

This type of setup will increase testing time but should be used on the problem areas. A full compliance test configuration will mean your pre-compliance testing will provide a greater degree of visibility into the EMI profile of your design.

On many instruments, you can also store cable and antenna correction factors that will allow you to see the true signal, without the added errors from the setup.

The next step in radiated testing includes using antennas in place of the near field probes, a rotating platform for the equipment under test (EUT), and can include a special room that minimises environmental factors (semi-anechoic). These setups are beyond the scope of this document, but there are references at the end that provide good references for the details of the setup.

Figure 4: Compliance Setup for Radiated Emissions Testing


Measuring Conducted EMI
Conducted EMI testing requires analysing the RF energy that is coupled from the instrument or test circuit to the main power line it is connected to.

Like Radiated EMI, Conducted EMI is also measured using a spectrum analyser, but it also requires a transient limiter and a Linear Impedance Stabilisation Network (LISN). A LISN isolates the power mains from the equipment under test, isolates any noise generated by the EUT,  and couples the signals generated by the EUT to the spectrum analyser.
 

Figure 3: Standard Conducted Emissions pre-compliance setup via a transient limiter. Note: be sure to choose a limiter suited to your own system and testing regime and do your own checks first: transients can easily exceed the input rating of your analyser (e.g. at switch-on - it is good practice to avoid switching on the LISN with the analyser connected) A modern DSA's internal level indicator is not intended to warn against fast transients and internal attenuators do not provide greater input circuit protection - if in doubt do your own measurements first to ensure you always operate within the stated input limit envelope.

As with emissions testing, the best start is a scan over the frequency range of interest using the peak detector on the spectrum analyser. Then, performing a quasi-Peak scan using the EMI filter for the problem areas. This will minimise test time while maintaining a high degree of confidence in your test.

Summary
EMC Compliance testing is mandatory for the majority of electronic products that are slated for sale throughout the world. For the cost of 1 day of compliance testing, you can have a pre-compliance setup that you can use to continually monitor and improve your design. This will help speed product development and  save the company money.

Was this article helpful?

Comments:
 

Related Articles
 > Is there an attenuator built-in on RIGOL's DSA815?
 > Is there an attenuator built-in on RIGOL's DSA1030?
 > Are there any independent reviews of RIGOL Spectrum Analysers like DSA815 in the UK?
 > How do I add a horizontal Display Line to my DSA screen?
 > How to reduce or lower noise floor? (DSA815 and DSA1030)
 > How to get the specified DANL of -135dBm? (typ, +6dB f/GHz)
 > How do I check my DSA is working properly?
 > How do radio amateurs use RIGOL's DSA815? And for what kinds of applications?
 > I just received mine… how can I tell if it's working OK?
 > Can I get help on using complex functions of my instrument?
 > Can I get help with programming my instrument?
 > Will I get a UK phone number for the team who sold to me?
 > How do I achieve the lowest DANL with my spectrum analyser?
 > How to measure an RF Amplifier using a DSA-800 Series Spectrum Analyzer
 > How do I measure an RF Bandpass filter
 > How do I connect my instrument via my LAN to UltraSigma PC Software?
 > After receiving my instrument(s), what should I do with my RIGOL Warranty Card?
 > Can I get Rigol firmware updates?
 > When installing Ultra Sigma I get a message PC telling me to install NIVISA / VISA Runtime engine. What am i doing wrong?
 > What about RIGOL battery and portable use? Can you run RIGOL Scopes and Spectrum Analysers from Battery or portable UPS Power Supplies?
 > USB stick / flash drive / memory sticks must fully compliant with USB2.0* standard compliant w/Win7 - Info on good sticks, esp. earlier models:
 > RIGOL 3GHz Spectrum Analyser - Applications insights from RIGOL-UK's customers.
 > Getting started with Rigol PC Software Tools
 > AMK: What do you get with the DSA Advanced Measurement Kit ? (DSA1000-AMK)
 > test
 > How do i measure an RF Attenuator?
 > Using a Rigol product key to generate a Rigol software licence code.
 > Rigol PC Software Summary
 > Rigol PC Software Versions
 > Can I get IVI drivers for my Rigol test equipment?